Description

The PHDY.2005 controls the PHDX.2005 unit. It yields power for both PMTs and control voltage for PMTs. The signals from PMTs are conditioned : amplified, compared (low and high tresholds), converted to digital level signals and counted. Three counters are available for detection counting : Channel A, Channel B and Channel (A & B). All counters are resetable.

PHDX.2005 and PHDY.2005 are devices primarily designed for photoncounting experiments. They show the quantum behaviour of photons. They clearly demonstrate the influence of various parameters on photon counting statistics (input amplification, dark current, treshold setup, PMT sensitivity and PMT control voltage, input flux, etc.). These devices are suitable for testing of parameters of PMTs (photon multiplier tubes).
See also PHDX.2005

Parameters

PMT control voltage

0.25 to 0.9 V

PMT DC power

12V

Amplification

1×, 10×

Digital output

5V, 20 µs

Treshold level – low

0 to 5 V

Treshold level – high

0 to 5 V

Max counter rate

10 kHz

Input and output signal connectors

BNC

PMT power connectors

M8

Power

230 VAC, 50/60 Hz

Power line system

1/N/PE

Dimensions

367 × 132 × 250 mm

Weight

3 kg

Related products